Product Details

The Bently Nevada 177313-02-01, also cataloged as the 177313 Electric Driven TK-3e, operates as a dedicated hardware component for calibration and field verification of proximity transducers within Bently Nevada Machinery Monitoring Systems.

Suffix Breakdown & Model Matrix

Option Field Code Configuration Detail
Scale Units Option 02 Metric calibration scale
Power Cord Type Option 01 American standard power cord

Hardware Specifications

Parameter Specification
Model 177313-02-01
Brand Bently Nevada
Origin USA
Weight 8 kg shipping weight
Dimensions 19.5 cm x 29.9 cm x 24.8 cm (H x W x D)
Operating Temp 0 deg C to 54 deg C (32 deg F to 130 deg F)
Power Consumption Line-powered electric motor drive mechanism
Storage Temp -18 deg C to 65 deg C (0 deg F to 150 deg F)
Environmental Protection Designed to meet IP54 for dust and water exposure (closed)
Operating Humidity 95% non-condensing humidity

Transducer Calibration and Gap Voltage Validation

The 177313-02-01 electric-driven calibration instrument simulates mechanical shaft motion to verify eddy-current probe scaling and Proximitor sensor response across defined displacement ranges. Utilizing an electric motor-driven wobble plate assembly, the unit maintains mechanical alignment to test target displacement against probe output voltages. This calibration setup enables field verification of gap voltage targets (-10 VDC nominal setting) and linear scale factor calculations without disturbing installed rotor dynamic monitoring channels or dismantling machine bearing housings.

Frequently Asked Questions

Q: How does the metric scale unit option affect displacement measurements?

A: Option 02 equips the micrometer assembly with metric increments, providing direct mechanical displacement reference readings in millimeters during transducer calibration routines.

Q: What protection rating applies when the instrument enclosure is closed?

A: The closed housing assembly meets IP54 dust and splash water ingress protection standards for transport across field environments.

Field Installation Guidelines

  1. Position the instrument on a flat, stable surface within reach of an AC power outlet matching the option 01 American cord configuration.
  2. Inspect the micrometer spindle, target plate, and probe mounting adapter to confirm surfaces are free of debris or mechanical burrs.
  3. Secure the proximity probe into the probe holder, ensuring the sensor tip remains parallel to the target disc surface.
  4. Connect the proximity probe extension cable to its corresponding Proximitor sensor and establish multimeter connections across the DC voltage output terminals.
  5. Rotate the micrometer dial to verify physical gap zeroing before adjusting spindle displacement to validate probe linearity across the full operating range.

Additional Information

  • 100% Genuine Parts: All products are original and authentic, ensuring reliable industrial performance.
  • 30-Day Refund Guarantee: Return any in-stock item within 30 days in original, unopened packaging for a full refund (excluding shipping and fees).
  • 12-Month Warranty: Covers defects in materials or workmanship; excludes misuse, normal wear, or unauthorized modifications.
  • Worldwide Shipping: We ship via USPS, UPS, FedEx, and DHL. Delivery times vary by country and may be subject to customs or import fees.
  • Support & Contact: Technical and warranty assistance is available anytime. Contact us here: Contact.
  • Purchase Guidance: Check product specifications and compatibility carefully before ordering to ensure proper application.




Recently Viewed Products

Tech & Buying Guide

Technical Insights, Installation Guides, and Buying Tips
Why Programmable Logic Controllers Drop into Stop Mode: An Expert Analysis

Why Programmable Logic Controllers Drop into Stop Mode: An Expert Analysis

Industrial automation systems rely heavily on Programmable Logic Controllers (PLCs) to maintain uninterrupted production. When a PLC unexpectedly shifts from Run to Stop mode, whole manufacturing lines stall.

As control systems evolve across modern factory automation, understanding the root causes of these interruptions helps engineers minimize costly downtime. This technical guide breaks down why PLCs trip into Stop mode and provides practical strategies to mitigate system faults.

Read more
Understanding Root Causes of Unplanned PLC Stop Mode in Industrial Automation

Understanding Root Causes of Unplanned PLC Stop Mode in Industrial Automation

Modern Programmable Logic Controllers manage execution through distinct operational states: Run, Stop, Start-Up, and Hold. During power-up, a controller initializes memory registers within Stop mode before briefly transitioning to Start-Up mode. During this phase, internal self-diagnostics evaluate system integrity across hardware and user programs. If the controller detects no critical faults, it enters Run mode to execute the control program continuously. However, any severe system anomaly forces the CPU back into Stop mode immediately to prevent erratic machine behavior across the plant floor.

Read more
Human-Machine Interface Design: Essential Guide to HMI Selection, Visualization, and Control System Integration

Human-Machine Interface Design: Essential Guide to HMI Selection, Visualization, and Control System Integration

Human-Machine Interfaces (HMIs) serve as the primary operational window into complex industrial automation systems. Modern HMIs convert raw machine data into clear graphical displays, enabling operators to oversee entire manufacturing processes in real time. Instead of relying on physical control panels full of hardwired switches, engineers program interactive screens using intuitive graphics, dynamic color coding, and responsive touch interfaces. Over my 15 years in control systems engineering, I have observed HMIs transform from basic display screens into sophisticated command centers that drive operational efficiency across DCS and PLC environments.